The 2.83L online monitoring particle counter complies with the standards of online dynamic monitoring and meets the requirements of ISO14644-1:2015 and GIP. 2.83L particle counter (0.1CFM) is used to meet different cleanroom applications for routine remote cleanroom online real-time monitoring. The interactive software will guide users to complete dynamic monitoring of clean room clean environment in accordance with ISO14644-2015 and GIP appendix。
Advantage
Function
Apply
The cleanroom monitoring results can be displayed on a terminal display via RS485, WiFi, LORA, or Ethernet. The device features a robust, chemically resistant (SUS316L) casing that is lightweight and portable. Its streamlined design minimizes particle traps, making the equipment easy to clean. It uses a 24V DC power supply to power the components. An optional external wireless adapter facilitates wire less network communication, simplifying network system integration. The system connects to third-party monitoring systems via Modbus TCP/IP or Modbus RTU protocols for remote management of all particle counters within the system.
• Through the combination of online monitoring software, real-time monitoring, historical curve,alarm record, data record, audit tracking and other functions can be realized.
• 24V DC power supply for continuous use.
• Store samples to reduce operator error.
• Use data management software to archive long term data and support long-term data storage.
• Supports WiFi and LORA communication for easy data transmission and device control (optional). CSV files can be downloaded for easy data integration.
• Interactive software guides users to complete cleanroom environment monitoring according to ISO 14644-1 and GMP standards.
• Alarm value can be set to facilitate fault identification in cleanroom.
• Modbus TCP/IP or Modbus RTU communication protocol can communicate with third-party monitoring systems.
• Built-in low noise power source, providing a comfortable working environment.
• Download to comply with 21 CFR-11 rules, user configuration privileges, access rights.
• Compatible with conventional cleanroom environment and other clean environment monitoring
• Complies with ISO 14644-1:2015 standard
• Analyze and generate reports in accordance with ISO 14644-1:2015, GMP
• Particle trend analysis
• Statistical process control
• Particle drift accident diagnosis
| Model | SX-3014E(no display) | SX-3014UE | SX-5014P(no display) | SX-5014UP |
| Particle size channel | 0.3μm,0.5μm,1.0μm,5.0μm | 0.5μm,1.0μm,5.0μm,10.0μm | ||
| Sampling flow | 2.83L/min ±5% (0.1 CFM) | |||
| Laser light source | Imported laser diodes (continuous life up to 100,000 hours) | |||
| Vacuum power source | The continuous working flow of the DC fan is stable, energy saving and environmental protection, and it is efficient and low noise | |||
| Verification criteria | JJF1190-2008,GB/T6167-2007 | |||
| Relative error of repeatability | ≤±10%FS | |||
| Particle size distribution error | ≤±30% | |||
| Particle concentration indication error | ≤±30%FS | |||
| Self-cleaning time | ≤10min (counting is zero for 3 consecutive times within 10 minutes) | |||
| Sampling time | Set the range: 1 to 65,535 seconds | |||
| Sampling delay | Set the range: 1~255 seconds | |||
| Number of samples | Range: 1 to 65,535 | |||
| Channel verification | Each channel DAC can be configured and independently verified | |||
| Sampling concentration | 35,000 particles /L | |||
| Counting mode | Cumulative number, concentration | |||
| Data memory capacity | 1~20,000 sets of measurement data, 4G data storage space, can be queried | |||
| Data service | PC terminal software | |||
| Communication mode | RS485: Modbus RTU protocol, Ethernet: Modbus TCP/IP protocol | |||
| Other optional interfaces | WIFI/LORA is optional and supports Modbus protocol | |||
| Cleanliness level judgment standard | ISO 14644-1, GMP dynamic, GMP static | |||
| Power supply | DC24V 1A | |||
| Other accessories | RS485 terminals, power adapter | |||
| Data communication interface | USB disk export, support Excel table | |||
| Three levels of authority | Three levels of permission, data storage and retrieval, printing | |||
| Environmental requirements | Usage environment: temperature 0~40 relative humidity 10~70% | |||
| storage environment: temperature-30 ~45 relative humidity 0~90% | ||||
| Outline dimension | Length 190 x width 80 x height 126.5 (mm) excluding interface | |||
| Exterior material | SUS316L stainless steel | |||
| weight | 1.6Kg | |||
| Quality Assurance | The warranty is 1 year, and optional extended warranty service can be selected | |||