The 2.83L online particle counter meets the standards for
online dynamic monitoring, complying with ISO 14644-1:2015 and GMP requirements. This product is designed for various cleanroom applications and is used for
routine remote real-time monitoring of cleanrooms. The interactive software guides users to perform dynamic monitoring of the clean environment in cleanrooms according to ISO 14644-1:2015 and GMP annexes.
The cleanroom monitoring results can be displayed on a
terminal display via RS485, WiFi, LORA, or Ethernet. The device features a robust, chemically resistant (SUS316L)
casing that is lightweight and portable. Its streamlined design minimizes particle traps, making the equipment easy
to clean. It uses a 24V DC power supply to power the components. An optional external wireless adapter facilitates wire
less network communication, simplifying network system integration. The system connects to third-party monitoring
systems via Modbus TCP/IP or Modbus RTU protocols for remote management of all particle counters within the
system.
Advantage
• Provide a complete verification manual to facilitate compliance with pharmaceutical/electronic specifications.
• Through the combination of online monitoring software, real-time monitoring, historical curve,alarm record, data record, audit tracking and other
functions can be realized.
• 24V DC power supply for continuous use.
• Store samples to reduce operator error.
• Use data management software to archive long term
data and support long-term data storage.
• Supports WiFi and LORA communication for easy
data transmission and device control (optional). CSV
files can be downloaded for easy data integration.
• Interactive software guides users to complete cleanroom environment monitoring according to ISO 14644-1 and GMP standards.
• Alarm value can be set to facilitate fault identification
in cleanroom.
• Modbus TCP/IP or Modbus RTU communication protocol can communicate with third-party monitoring systems.
• Built-in low noise power source, providing a comfortable working environment.
• Download to comply with 21 CFR-11 rules, user configuration privileges, access rights.
Function
• Provide 2.83L/min(0.1CFM) traffic collection
• Compatible with conventional cleanroom environment and other clean environment monitoring
• Complies with ISO 14644-1:2015 standard
• Analyze and generate reports in accordance with ISO 14644-1:2015, GMP
Apply
• Cleanroom clean environment testing
• Particle trend analysis
• Statistical process control
• Particle drift accident diagnosis
Technical Specifications
Model
|
SX-3014E(no display)
|
SX-3014UE
|
SX-5014P(no display)
|
SX-5014UP
|
Particle size channel
|
0.3μm,0.5μm,1.0μm,5.0μm
|
0.5μm,1.0μm,5.0μm,10.0μm
|
Sampling flow
|
2.83L/min ±5% (0.1 CFM)
|
Laser light source
|
Imported laser diodes (continuous life up to 100,000 hours)
|
Vacuum power source
|
The continuous working flow of the DC fan is stable, energy saving and environmental protection, and it is efficient and low noise
|
Verification criteria
|
JJF1190-2008,GB/T6167-2007
|
Relative error of repeatability
|
≤±10%FS
|
Particle size distribution error
|
≤±30%
|
Particle concentration indication error
|
≤±30%FS
|
Self-cleaning time
|
≤10min (counting is zero for 3 consecutive times within 10 minutes)
|
Sampling time
|
Set the range: 1 to 65,535 seconds
|
Sampling delay
|
Set the range: 1~255 seconds
|
Number of samples
|
Range: 1 to 65,535
|
Channel verification
|
Each channel DAC can be configured and independently verified
|
Sampling concentration
|
35,000 particles /L
|
Counting mode
|
Cumulative number, concentration
|
Data memory capacity
|
1~20,000 sets of measurement data, 4G data storage space, can be queried
|
Data service
|
PC terminal software
|
Communication mode
|
RS485: Modbus RTU protocol, Ethernet: Modbus TCP/IP protocol
|
Other optional interfaces
|
WIFI/LORA is optional and supports Modbus protocol
|
Cleanliness level judgment standard
|
ISO 14644-1, GMP dynamic, GMP static
|
Power supply
|
DC24V 1A
|
Other accessories
|
RS485 terminals, power adapter
|
Data communication interface
|
USB disk export, support Excel table
|
Three levels of authority
|
Three levels of permission, data storage and retrieval, printing
|
Environmental requirements
|
Usage environment: temperature 0~40 relative humidity 10~70%
|
storage environment: temperature-30 ~45 relative humidity 0~90%
|
Outline dimension
|
Length 190 x width 80 x height 126.5 (mm) excluding interface
|
Exterior material
|
SUS316L stainless steel
|
weight
|
1.6Kg
|
Quality Assurance
|
The warranty is 1 year, and optional extended warranty service can be selected
|